What Changed

The Medicines and Healthcare products Regulatory Agency (MHRA) issued a Class 4 Medicines Defect Notification on 25 November 2026 for Takeda UK Limited’s PROSTAP PD DCS 1.88 mg. The leuprorelin acetate product, supplied as powder and solvent for prolonged-release suspension for injection in a pre-filled syringe, must only be administered subcutaneously.

Section 4.2 of the Summary of Product Characteristics (SmPC) and the outer carton specify the correct route. However, Step 5 of the administration instructions in the SmPC and Healthcare Professionals’ User Leaflet incorrectly permits intramuscular administration. The immediate syringe label contains the same error.

Who Is Affected

The notification concerns prescribers, dispensers and administering healthcare professionals, particularly treating paediatric endocrinology teams, alongside patients and carers. It covers batches 523373 and 518961, each supplied in packs containing one pre-filled syringe.

Healthcare professionals should continue prescribing, dispensing and administering the medicine where clinically appropriate, using only the subcutaneous route. Patients should not stop or change treatment without advice from their specialist or another healthcare professional.

Regulatory and Technical Implications

The issue is a labelling defect. No manufacturing, formulation, identity, strength or purity defect has been identified. No confirmed incorrect-route administrations or associated adverse clinical outcomes have been identified for the UK presentation.

Available evidence does not indicate serious systemic harm or clinically meaningful loss of efficacy. Takeda classified the expected severity of harm as minor, mainly involving temporary, manageable injection-site reactions. However, direct evidence comparing the pharmacokinetic and pharmacodynamic effects of subcutaneous and intramuscular administration is unavailable.

Treating paediatric endocrinology teams should review administration records. Where intramuscular administration may have occurred, appropriate clinical follow-up should be considered, including confirmation of continued suppression of pubertal progression.

Manufacturer Actions

Takeda is progressing corrective regulatory action to ensure product information and labelling consistently state the approved subcutaneous route.

To support UK supply, the MHRA, in discussion with the Department of Health and Social Care, considers that currently packed batches available for distribution will continue to be distributed until replacement stock becomes available.

Confirmed Dates

The notification was published and issued on 25 November 2026. Batch 523373 was first distributed in October 2025 and expires in September 2027. Batch 518961 was first distributed in December 2024 and expired in March 2026. The expired batch remains within the notification and associated patient follow-up scope because it carried the same defect.

Replacement stock is anticipated in March/April 2027. This is an expected availability window, not a confirmed delivery date.

Official Source

MHRA Class 4 Medicines Defect Notification EL(26)A/52; DMRC reference: DMRC-41456133. Suspected adverse reactions should be reported through the MHRA Yellow Card scheme.

Https://www.gov.uk/drug-device-alerts/class-4-medicines-defect-notification-takeda-uk-limited-prostap-pd-dcs-1-dot-88-mg-powder-and-solvent-for-prolonged-release-suspension-for-injection-in-pre-filled-syringe-el-26-a-slash-52

Source and related services